Vulnerability Monitor

The vendors, products, and vulnerabilities you care about

CVE-2019-18241


In Philips IntelliBridge EC40 and EC80, IntelliBridge EC40 Hub all versions, and IntelliBridge EC80 Hub all versions, the SSH server running on the affected products is configured to allow weak ciphers. This could enable an unauthorized attacker with access to the network to capture and replay the session and gain unauthorized access to the EC40/80 hub.


Published

2019-11-26T00:15:11.717

Last Modified

2024-11-21T04:32:54.213

Status

Modified

Source

[email protected]

Severity

CVSSv3.1: 6.5 (MEDIUM)

CVSSv2 Vector

AV:A/AC:L/Au:N/C:P/I:N/A:N

  • Access Vector: ADJACENT_NETWORK
  • Access Complexity: LOW
  • Authentication: NONE
  • Confidentiality Impact: PARTIAL
  • Integrity Impact: NONE
  • Availability Impact: NONE
Exploitability Score

6.5

Impact Score

2.9

Weaknesses
  • Type: Secondary
    CWE-326
  • Type: Primary
    CWE-326

Affected Vendors & Products
Type Vendor Product Version/Range Vulnerable?
Operating System philips intellibridge_ec40_firmware * Yes
Hardware philips intellibridge_ec40 - No
Operating System philips intellibridge_ec80_firmware * Yes
Hardware philips intellibridge_ec80 - No

References